Modell SET-206
|
The Model 206 industrial pressure transducer is one of the most rugged and reliable sensors available. These transducers prove to be superior to competitive brands and technologies in application field testing.
The capacitive design exhibits resistance to environmental effects such as shock, vibration, temperature and EMI/ RFI. The Model 206 also meets NEMA-4 and IP65 environmental protection ratings.
With a highly accurate stainless steel capacitive sensor design and custom ASIC based circuit, the Model 206 maintains a high level of performance and accommodates a variety of pressure fittings and electrical connector options.
Spezifikationen | |||
---|---|---|---|
Performance Data | |||
Accuracy RSS (at constant temperature) | ±0.13% FS | [1] | |
Non-Linearity, (BFSL) 25 psig range | ±0.1% FS ±0.2% FS | [2] | |
Hysteresis | 0.08% FS | ||
Non-Repeatability | 0.02% FS | ||
Response Time | 5 milliseconds | ||
Long Term Stability | 0.5% FS/1 YR | ||
Thermal Effects | |||
Compensated Range | -20 to +80°C | ||
Zero Shift | 0.9 %FS/50°C | ||
Span Shift | 1.4 %FS/50°C | ||
Physical Description | |||
Pressure Fittings | See Spec Sheet | ||
Vent | Through cable or terminal | ||
Electrical Connection | 2 ft. multiconductor cable | ||
Case | Stainless Steel | ||
Zero/Span Adjustments | Top External Access | ||
Weight (approx.) | 6 oz | ||
Environmental Data | |||
Operating/Storage Temperature | -40 to +85°C | [4] | |
Temperature | -40 to +85°C | ||
Acceleration | 10g Maximum | [5] | |
Shock | 200g Operating | [6] | |
Vibration | 20g 50-2000 Hz | [7] | |
Pressure Media | Gases or liquids compatible with 17-4 PH Stainless Steel. | [3] | |
Electrical Data (Voltage) | |||
Excitation/Output | 12 to 28 VDC Reverse Excitation Protected | ||
Power Consumption | <0.15 watts (approx. 5mA at 24VDC) | ||
Output | 0.1 to 5.1 VDC | [8][9] | |
Output Impedence | 100 ohms | ||
Circuit | 3-Wire (Exc, Out, Com) | ||
Electrical Data (Current) | |||
Circuit | 2-Wire | ||
Output | 4 to 20 mA | [10][11] | |
External Load | 0 to 800 ohms | ||
Min. Supply Voltage (VDC) | 9 +0.02 x (Resistance of receiver plus line) | ||
Max. Supply Voltage (VDC) | 30 + 0.004 x (Resistance of receiver plus line) |
Hinweise | |||
---|---|---|---|
[1] | RSS of Non-Linearity, Non-Repeatability and Hysteresis | ||
[2] | 25 psig range accuracy is ±0.22% of Full Scale output | ||
[3] | Hydrogen not recommended for use with 17-4 PH stainless steel. | ||
[4] | The high temperature limit of the cable is 200°F (95°C) | ||
[5] | Shift in output reading <0.05 psi/g typical; pressure port axis only | ||
[6] | Mil-Std. 202, Method 213B, Cond. C | ||
[7] | Mil-Std. 202, Method 204, Cond. C | ||
[8] | Calibrated into a 50K ohm load, operable into a 5000 ohm load or greater | ||
[9] | Zero output factory set to w/in ±25mV. Span (FS) output factory set to w/in ±50mV. | ||
[10] | Calibrated at factory with a 24VDC loop supply voltage and 250ohm load. | ||